If a depleted battery is not able to be disposed of properly, label the
battery in some manner to prevent it from being reused.
To Install the Battery
1.
With the label on the battery facing the Survivalink AED battery
compartment, insert the battery as shown in the drawing.
2.
Push the latched end of the battery firmly into the Survivalink AED,
as shown in the drawing, until the battery snaps into place—you will
hear a loud click. The exposed side of the battery should be flush with
the outside of the Survivalink AED case.
If the battery is not properly installed, the Survivalink AED will not
operate.
Activating the Survivalink AED Self-Tests
The “Status” indicator on the Survivalink AED handle will only turn
GREEN (RescueReady) after the self-test is complete. To activate a self-
test:
1.
Open the lid of the Survivalink AED and the following should occur:
• The indicator lights on the diagnostic panel will toggle sequentially
On/Off while the Survivalink AED performs its daily self-tests
• After the Survivalink AED completes the tests, you will hear the
voice prompt,
“Place Electrodes” -
The status indicator on the
Survivalink AED handle will switch to GREEN
2.
Close the lid—the Survivalink AED is RescueReady.