Section 4: Rescue
P
age 48
©2002 Cardiac Science, Inc.
300282-002 Rev. E0
Post Rescue Process
After transferring the patient to ALS personnel, do the following to
prepare the Survivalink AED for the next rescue:
1.
Connect a new pair of electrodes to the Survivalink AED.
2.
Check the expiration date on the electrode package for expiration.
3.
Place one electrode expiration date sticker on the outside of
Survivalink AED where it can be viewed without opening the lid.
4.
Place the new electrodes in the storage space.
5.
Verify that the “Replace” indicator is not lit. If it is, replace the
battery.
6.
Close the lid.
7.
Verify that the “Status” indicator on the Survivalink AED handle is
GREEN.
8.
Retrieve the rescue data stored in the internal memory of the
Survivalink AED or from a Rescue Data Card by using RescueLink
installed on a PC. (Complete this step before the next rescue to
prevent loss of rescue data).
After retrieving the rescue data, erase the internal memory of the
Survivalink AED or the Rescue Data Card before preparing for a new
rescue as outlined in the RescueLink User Manual.
9.
Insert a blank Rescue Data Card into the Survivalink AED card slot
(Model 9110 or 9210 only).