300282-002 Rev. E0
©2002 Cardiac Science, Inc.
Page 73
Section 7: Technical Data
Emissions
Immunity
Field
Models
Standard or Compliance
E-M
9100/9110
9200/9210
EN 55011/C.I.S.P.R. 11, Group 1, Category B
RTCA/DO-160D, Section 21, Category L
(Category B during charging)
RTCA/DO-199, Section 6.2.2
Magnetic
9100/9110
9200/9210
AAMI DF39, < 0.5mT on surface, except for
within 5cm of the lid magnet and the speaker
RTCA/DO-160D, Section 15, Category Z
9200/9210
RTCA/DO-199, Section 6.2.1 during analysis
only
Field
Models
Standard or Compliance
E-M
9100/9110
IEC 801-3 (1984), Level 3
9200/9210
EN 61000-4-3, Level 3
Magnetic
9100/9110
MIL-STD-462D, Method RS101
9200/9210
EN 61000-4-8, 80A/m for 47.5Hz - 1320Hz
ESD
9100/9110
IEC 801-2 (1991), Level 2
9200/9210
EN 61000-4-2, Level 2