Section 2: Introduction
The European CPR Guidelines or Universal ALS (European Protocol)
recommends up to 3 minutes of CPR if the first analysis decision is non-
shockable or following two consecutive non-shockable analysis decisions.
The three (3) defibrillation shocks are delivered in a pre-programmed
sequence of escalating monophasic, or biphasic energies.
The Survivalink AED Models 9100/9110 first defibrillation shock is set
at 200J. The second defibrillation shock is programmable for either 200J
or 300J. The third defibrillation shock and all subsequent shocks are set
at the maximum energy which is 360J.
The Survivalink AED Models 9200/9210 first defibrillation shock is set
at a low current
3
setting. The second defibrillation shock is programmable
for either low current or high current. The third defibrillation shock and
all subsequent shocks are set at the high current setting.
3. FirstSave Models 9200/9210: The low current and high current shocks are variable energy. The actual energy is
determined by the patient’s impedance.