CK and CN filters correspond to the
electric filters values used by the cell.
Two possible commands from this page:
a) Autozero :
This key is to automatically launch the autozero procedure.
b) Error assessment from opacity values: « TEST »
It is about an automatic check done by electric filters around 50% and allowing to verify measures
accuracy. The error is considered acceptable if not over 5%. Press “TEST” to launch this operation.
Before launching the test, proceed an autozero if opacity values are not equal to 0.
A window will appear showing the result. If this last shows “smoke-meter not conform”, clean the
cell and launch the test again.
V.5.2.Secured access :
This menu is exclusively accessible by the manufacturer, CAPELEC, as it gives access to secured
parameters.