V.4.1.Time settings :
The stand-by allows to stop
pumps after a test.
This time can be set
between 7 and 30 minutes.
The temporisation time
allows adding additional time
before starting a gas or smoke test.
V.4.2.Digits CO:
The DIGIT CO submenu allows selection of the number of figures after the decimal that you want
to have when CO is displayed
V.4.3.O2 sensor status:
This window provides information on the oxygen sensor status. The machine performs a
diagnostic on the sensor and gives its results of the sensor wear condition. Result is graduated
from 0 to 100%, 100% corresponding to the most satisfactory status.
V.4.4.Leak test
The leak test is a procedure allowing the detecting of any seal problems that might occur in the
pneumatic circuit. It is imperative to carry out this test at least once every three months. It is also
preferable that it be carried out at least once per month by the user. The presence of a leak
involves immediate maintenance work.
This test is guided throughout. The operator must attentively follow the instructions given on the
screen.
Test upstream of the pump:
The following message appears: “Plug the inlet”. Blocking of the inlet is automatically detected by
the analyzer. The pumps then shut off and the following message appears: “Test in progress.” It is
Содержание EOBD
Страница 2: ......
Страница 7: ...23 04 2019 Rev 3 4 I 2 Overview I 2 1 The Station ...
Страница 10: ...23 04 2019 Rev 3 7 I 2 4 Smoke meter Exhaust probe ...
Страница 13: ...23 04 2019 Rev 3 10 CAP4XXX RPM via the EOBD Allows to measure the RPM from the vehicle embedded computer ...
Страница 23: ...23 04 2019 Rev 3 20 II 1 3 Gas flow circuit diagram ...
Страница 63: ...23 04 2019 Rev 3 60 ...