II.3.
Technical features
Preheating duration: < 9 minutes at 0°C (1 minute minimum)
Response time: 13 seconds for HC, CO, CO2, 28 seconds for oxygen (transition from 20.9%
to 0.1% for a gas with 0% O2)
Nominal pump delivery: 6 l/min.
Minimum pump delivery: 3.5 l/min.
Air pressure variation: automatic correction by integrated absolute pressure sensor
Zero point and sensitivity: automatic compensation
Supply 115-230V / 1.5A ; 47Hz à 63Hz ; Fuses T2AL 250V.
o
Gas table:
Specified ranget
Display range
Resolution
CO
0 to 5 % vol
0 to 15 % vol
0,01 % vol or 0,001 % vol
CO
2
0 to 16 % vol
0 to 20 % vol
0,1 % vol
HC (C
6
H
14
)
0 to 2000 ppm vol
0 to 10000 ppm vol
1 ppm vol [0;2000 ppm]
and 10 ppm above
O
2
0 to 21 % vol
0 to 25 % vol
0,01% vol [0 ; 4 % vol]
0,1% vol above
λ
0,8 to 1,2
0,8 to 1,2
0,001
o
Measurement ranges:
NOx =
0 à 5000 ppm
Engine speed =
0 à 9999 RPM
Oil temperature =
-5 à 150 °C
Corrected CO =
0 à 10 %
o
Accuracy:
Engine speed =
+- 10 RPM
Oil temperature =
+- 1 °C
Corrected CO =
0,03 %
o
Other resolutions:
NOx =
1 ppm vol.
Engine speed =
1 tours/min
Oil temperature =
1 °C
Содержание EOBD
Страница 2: ......
Страница 7: ...23 04 2019 Rev 3 4 I 2 Overview I 2 1 The Station ...
Страница 10: ...23 04 2019 Rev 3 7 I 2 4 Smoke meter Exhaust probe ...
Страница 13: ...23 04 2019 Rev 3 10 CAP4XXX RPM via the EOBD Allows to measure the RPM from the vehicle embedded computer ...
Страница 23: ...23 04 2019 Rev 3 20 II 1 3 Gas flow circuit diagram ...
Страница 63: ...23 04 2019 Rev 3 60 ...