1. General
1
1 General
CXDI-60G/60C represents a compact cassette model that enhances user convenience in cassette-
based digital radiography. It is used in combination with Control computer (CXDI-C3, CXDI-C3S,
FC-24VE, FC-E21A or the general computers with equal performance).
The sensor cable is organized into two parts. The one is a sensor cable S150-60 (imaging unit side)
and the other is a sensor cable P630 (power box side). Two cables are connected with the connectors.
It adopts the detachable connector, which is compatible with the CXDI-55G/55C, allowing the use
of two sizes portable flat panel detector.
The vertical scanning drive (Drv-IC) and the data read out (Amp-IC) are located in one side of
LANMIT. Image data is read out from one direction.
The appearance of the CXDI-60G is almost the same as that of the CXDI-60C, but their fluorescence
substances that convert X-ray into visible light are different.
CXDI-60G
CXDI Software
Ver.7.0 and later
Serial number of CXDI-60G imaging unit
100001 and later
Serial number of CXDI-60G imaging unit (non-
logo model)
300001 and later
CXDI-60C
CXDI Software
Ver.7.2 and later
Serial number of CXDI-60C imaging unit
100001 and later
Notes on usage
From the view of risk management, guarantee is not made for the waterproofing for blood and
chemicals, hygienic safety in operating room, usage with a defibrillation device in ICU, outdoor
usage or application to animals. If the imaging unit needs to be used under such condition, the
system integrator should be responsible for the operation and understanding of the tolerance of the
product specification.
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...