5 Specifications
The CXDI-60G/60C (Imaging Unit/Power Box) is the Digital Cassette that has the mobility and
can be used on the optional angles.
(1)
Imaging Unit
This unit consists of the internal sensor, PWB-60Di, PCB-AD/REF, PWB-60LED and its
outer cover. The sensor unit converts the X-ray image to the electrical signal (O/E
Conversion) and after performing the A/D conversion, transfers its signal through the Power
Box with Ethernet cable to the Control computer.
Item
CXDI-31
CXDI-60G/60C
Remarks
Object Cassette
(built-in)
Cassette (built-in)/Mobile
Effective filming
range
226.00 x 288.00 mm
234.24 x 284.16 mm
Number of Pixels
2260 x 2280
(About 6,500,000 pixel)
1476 x 1790
(About 2,600,000 pixel)
Effective
Number of Pixels
2256 x 2878
1464 x 1776
Pixel pitch
100
µ
m x 100
µ
m 160
µ
m x 160
µ
m
Fluorescent
substance
GOS 60G:GOS/60C:CsI
Output
gradations
12bit (4,096 gradations)
A/D 14bit
Transfer method
Metal: Imaging Unit to E/O BOX
Optical fiber: E/O BOX to
Control computer (Through the
Multi Box)
Ethernet: Imaging Unit to Control
computer (Through the Power Box)
Imaging time
~ 1 sec.
Imaging cycle
15 sec.
Dimension
324(W) x 327(D) x 20.8(H) mm
344.4 (W) x 380* (D) x 22.5 (H) mm
Imaging Unit
coloring
Art gray
Cool white
Imaging Unit
mass
(w/o grid)
2.8Kg
(Except the cable)
3.3Kg
(Including the cable)
2.5Kg
(Except the cable)
2.8Kg
(including the cable)
Space between
surface where
patient gets in
contact (CFRP)
and sensor
surface (glass)
5.75mm Within
5.1±0.5mm.
Average (43.2kJ/h)
Remote SW OFF
36kJ/h or less
Heat emission
Max
162kJ/h
Max load mode
126kJ/h or less
1 image per 15 Sec.
Average (12W)
Remote SW OFF
10W or less
Power
Consumption
Max
45W
Max load mode
35W or less
(1 image per 15
Sec.)
Mechanical
strength
*
Load uniformly: 1.47kN(150kgf), 1min
Load locally: 0.98kN(100kgf),
φ
40mm, 1min
The imaging unit is put on the plain surface with the Sensor side (Detector) is
up.
Self-imposed rule
For CXDI-31, local
strength applies to
the center of the
sensor.
Control
computer
-CXDI-C2
-CXDI-C3
-FC-24VE/FC-E21A for CXDI Control
Station
-General computer with a performance
equal to FC-24VE/FC-E21A
-CXDI-C3/C3S
Ethernet card is
necessary to connect
CXDI-60G/60C with
CXDI-C3
Power Control
(ON/OFF)
Interface with a Control computer
None
(Power Box: Operation with Remote
switch manually)
Grid
attach/remove
detector
Yes
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...