6. Troubleshooting
13
6. How to Back-up and Recover the System
[Objective]
This document describes how to back up and recover the system.
[Technical Description]
There are two system backup methods as follow:
The recovery method required depends on the way the system crashes.
Backup method
A) Backup using floppy disks
B) Backup using a hard drive
6.1 Backup method
A) Backup method using floppy disks
The network settings configured in each installation site and the customized body part settings
are stored in each ini file.
All the ini files are stored on a floppy disk when starting up the control PC.
The system has a feature that stores the latest ini files on the floppy disk at the system startup
by updating the ini files with modifications users made while using the system.
B) Backup method using a hard drive
The control PC has no feature that mirrors all the files including the OS to other hard drives.
Therefore, to be ready for hard drive crashes, we recommend that you add another hard drive
when installing the system, in order to copy the software between the hard drives using Ghost
or Drive Copy, which are available on the market.
This should be conducted at the final installation stage (just before delivery to users).
As an alternative, you can also provide a hard drive that contains the OS (before activation)
and drivers in case of hard drive crashes.
6.2 Recovery method
6.2.1
When problems occur in a CXDI software ini file
The system can be recovered to the last environment status just before the system was used by
users when problems
*1
occur in d:\Ccr software other than the OS.
*1
When the setting data or files are damaged
In this case, the recovery can be made by overwriting ini files stored on the floppy disk to d:\ccr.
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...