4. Repair Guide
9
•
Grid foil direction:
User-selected. However, it must be parallel to the lengthwise or widthwise direction.
•
Grid lattice density:
40 lines/cm +/-5% (fluctuations in number of lines per unit: less than +/-1%)
•
Grid ratio/Focusing distance:
User-selected
•
Grid surface (for EU):
Biological safety (EN ISO 10993-1/5/10) must be taken into account for the surfaces that
can come into contact with the patient.
•
Grid mounting angle:
The tolerance must be less than 1 degree for the angle at which the grid is mounted vis-a-vis
the rows of sensor pixels.
However, an angle of less than 0.3 degree is preferred for optimal functioning of the grid
stripe reduction process.
Note: Mounting angle = Relative angle between the sensor pixel rows and grid stripe
•
Recommended grid manufacturers:
Mitaya Manufacturing, SMIT, JPI
•
Other
notes:
Other specifications not shown above should be compliant with JIS4910-2000.
3.3 Configuration
3.3.1 Grid unit configuration
The parts required for adhering the grid to the grid frame unit are shown in the table below.
NO.
PARTS NAME
PARTS NO. Q’TY
REMARKS
1
Grid
-
1
Obtained by the sales company
2
Grid frame unit
BM7-0061
1
Optional product
3
Grid spacer 4L (1.5 mm thick)
BA5-0322
2
4:1 grid ratio specification
(stamp mark 4)
4
Grid spacer 4S (1.5 mm thick)
BA5-0323
2
4:1 grid ratio specification
(stamp mark 4)
5
Grid spacer 8L (0.7 mm thick)
BA5-0325
2
8:1 grid ratio specification
(stamp mark 8)
6
Grid spacer 8S (0.7 mm thick)
BA5-0326
2
8:1 grid ratio specification
(stamp mark 8)
7
Grid spacer 0L (0.3 mm thick)
BA5-0328
2
10:1 grid ratio specification
(stamp mark 0)
8
Grid spacer 0S (0.3 mm thick)
BA5-0329
2
10:1 grid ratio specification
(stamp mark 0)
9
Double-sided tape (F9473PC made
by Sumitomo 3M, 0.255 mm thick,
11 mm wide)
- n
Used to adhere grid spacer
(same tape used for all spacers).
10
Double-sided tape (Y-4930 made by
Sumitomo 3M, 0.64 mm thick, 11.5
mm wide)
-
n
Used to adhere grid
* Parts indicated in blue are service parts.
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...