2. Installation
29
+5V(isolated)
Photo coupler
RX_REQ
RX_COM
Relay
PC1
RL1
Connector
J4
POWER BOX
X-ray I/F Cable
X-ray Generator
#2
#1
#4
#3
A2
A1
B2
B1
X-ray permission signal
X-ray request signal
*1:Impedance between A1-A2 < 100ƒ ¶
*2:Permitted Vold AC250Vmax/DC30Vmax
Permitted Current 10mA-2A
WHT
YEL
ORG
GRY
TLP281(GR/F)
NC2D-P-DC5V
The connection with CXDI-60G/60C and X-ray generator equipment
Connection conditions
1. The X-ray exposure signal line (including the switching function) must be insulated, and its total
impedance must be 100 ohms or less.
2. The maximum contact voltage of the X-ray exposure authorization signal line is AC 250V and
DC 30V, and its current ranges from 10mA to 2A.
Only the insulated secondary power supply can be connected.
3. Protective grounding for X-ray generator should be equipotent with the system.
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...