Page 1 of 2
CXDI-60G/60C Inspection Check Sheet
< I n s p e c t i o n I t e m s >
ITEM Category NO Item Characteristics values
Test means
Record
1 Grid
installation
It must be possible to
install and remove the
grid smoothly, and the
grid must securely lock
and unlock.
Touch
2
Grid/no grid
detection
Whether the grid is
present or not must be
properly detected.
Visual check
3 Sensor
cable
The cable must not be
flattened or twisted and
its covering must not be
damaged.
Visual check
4
Sensor cable
relay connector
The connector must be
connected securely,
and there must be no
play.
Visual check,
touch
Side panel right
Side panel left
Bottom panel
right
5 Shock
sensor
The sensor must detect
shocks and not turn red.
Bottom panel left
6 Firmware
version
-
Visual
check Ver.
7
PLD code version
-
Visual check
Ver.
1 Imaging
unit
8
Imaging unit IP
address
- Visual
check
1 Connector
The connector must be
connected securely.
Visual check,
touch
2 Power
switch
It must be possible to
set this switch to its ON
and OFF positions
properly.
Visual check,
touch
3 POWER
LED
The LED must light
when the power is on.
Visual check
The LED must not light
when the power is on.
2 Power
box
4 ERROR
LED
The LED must light
when the power is on
while the sensor cable
is disconnected from
the detachable
connector.
Visual check
1 Remote
switch
It must be possible to
set this switch to its ON
and OFF positions
properly.
Visual check
2
POWER LED
(Power Box must
be turned on.)
The LED must light
when the power is on.
Visual check
The LED must not light
when the power is on.
Visual check
3 Remote
switch
3
ERROR LED
POWER LED
(Power Box must
be turned on.)
The LED must light
when the power is on
while the sensor cable
is disconnected from
the detachable
connector.
Visual check
4 Control
PC
1
Connector
The connector must be
connected securely.
Visual check,
touch
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...