![Canon CXDI-55C Скачать руководство пользователя страница 110](http://html.mh-extra.com/html/canon/cxdi-55c/cxdi-55c_service-manual_61476110.webp)
2. Installation
94
7.7 Image Quality
(1) Purpose
This procedure is used to check the final image quality of the CXDI.
1) Resolution
check
Tools used
(1) Phantom
(2) High-resolution monitor or DICOM printer
Procedure
Procedure/Item Operation/Conditions
1. Adjust the distance between the focal point of the X-
ray tube and the CXDI imaging unit.
X cm: Differs according to the grid being used.
2. Perform calibration
SYSTEM > CALIBRATION
3. Perform self-test.
SYSTEM > SELF TEST
Must pass all the tests.
X-ray tube focal
point
X cm
Right angle at
the center
CXDI Sensor Unit
Self-test
[Positioning]
Position the X-ray tube and
CXDI Imaging Unit
Calibration
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...