![Canon CXDI-55C Скачать руководство пользователя страница 72](http://html.mh-extra.com/html/canon/cxdi-55c/cxdi-55c_service-manual_61476072.webp)
2. Installation
56
Parameter Meaning
Description
-v
This parameter dumps the
transfer log.
* This parameter is used
simply as “-v”.
* DUL and SRV are dumped.
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
* This parameter does not affect DICOM
data transfer.
-jn
This is the time to take
timeout.
* Sets the time to take timeout
in seconds.
* This parameter is to be changed when
taking timeout.
-k
level=0
1
2
3
* A variety of specifications
have since been needed in
conjunction with DICOM
modality LUT support.
If DICOM Modality LUT OD is enabled,
set appropriate options to suit each
output destination.
Argument: Values necessary for executing a function, subroutine, procedure, or other operation is
passed to them. Arguments are assigned to functions and subroutines when executed. For example,
the argument in f (x) is x.
PDU: Protocol data unit
The types of PDU’s include get-request, get-next-request, get-response, set-request, nd trap.
Note
For details about the “-v” parameter, see “Checking the Error Log”.
Note
The parameters “-v” and “-d” put the log on the console. Therefore, be sure to always erase
these parameters before operation by the user.
If –k:DICOM Modality LUT OD is enabled, set appropriate options to suit each output
destination.
level = 0: Do not delete (default when not specified)
1: Delete Window Center/Width
2: Delete Window Center/Width and Rescale Intercept/Slope/Type
3: Delete Rescale Intercept/Slope/Type (compatible with releases up to Ver.4.20)
Other than a loadable LUT or
γ
=1.0 has been specified with the output destination-
specific LUT function.
For storage: Specify -k3(or -k2).
If IMG Rescale Type = 0D is unidentifiable to storage, resulting in an error: Specify -k3.
(If only one storage is connected, simply set DICOM Modality LUT OD to "Disabled.")
The implementation allows Window Center/Width with Rescale Type = 0D specified to
be interpreted as "optical densityx1000."
For storage: Specify -k1.
The CXDI Window Center/ Width output value (implementation) is fixed at 2048/4096.)
Note
With AGFA impax Ver. 4.5.0, the specification of -k0 demonstrated a successful density-
intensity conversion. But because the corresponding text in the DICOM specifications
document is ambiguously written such that the status of implementation by other
manufacturers is unknown, please be advised to consult the storage manufacturer for each
connection destination or work out on a trial and error basis. Also note that an external
storage option specification (if (DICOM modality LUT OD is set to "Enabled," OD tags
<0028, 1052 - 1054> are assigned and Level: 0 is assumed) is not supported.
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...