1. General
6
Item
CXDI-31
CXDI-60G/60C
Remarks
Sensor cable
From imaging unit to E/O Box:
4.5m
Sensor cable (imaging unit side): 1.5m
Sensor cable (power box side): 6.3m
Count of
connected sensor
Up to four imaging units are
connectable to a single control
computer via multi box. For the same
type of imaging unit, the maximum of
connection number will be three.
Up to four Imaging Units are connectable to a
single control computer via Network switch.
For the same type of imaging unit, the
maximum of connection number will be three.
Each imaging unit requires a power box.
Network switch
should be procured
at each sales
company.
Scattered
radiation
backward block
sheet
Environment-
conscious
unleaded type
Lead cover
Mo sheet (0.3 mm thick)
Photo timer
Cannot be built in
Imaging
restriction
(Imaging
Prohibition)
When the internal temperature of
Imaging Unit is above 45 degree
Celsius, its state is changed to sleep
mode. And the Imaging prohibition
will be continued when the internal
temperature is below 44 degree
Celsius.
When the internal temperature of Imaging Unit
is above 49 degree Celsius, its state is changed
to sleep mode. And the Imaging prohibition
will be continued when the internal
temperature is below 48 degree Celsius.
User interface
Indication
Indication Busy Sensor Power
Color Two
colors
Color
Orange
*1
Green Blue
Imaging
unit is off
Off
Imaging
unit is off
Off Off
Off
Imaging
unit is on
Orange light will turn
on
Imaging
unit is on
-- Off On
Preparing
imaging
Green light will flash
*1
Preparing
imaging
--
Blinking
*2
On
Imaging
preparation
complete
Green light will turn on
Imaging
preparation
complete
-- On On
Error status
Green light will flash
*2
Error status
--
Blinking
*3
On
Communi-
cating
On --
On
Initialization
(when
startup)
--
Blinking
*4
On
Network not
set (when
startup)
--
Blinking
*5
On
*1: Turns on and off for 0.5 seconds
each
*2: Turns on and off twice for 0.5
seconds, then turns off for 0.5
seconds
*1: Green for China
*2: Turns on and off for 0.5 seconds each
*3: Turns on and off twice for 0.5 seconds,
then turns off for 0.5 seconds
*4: Fades in for 1 second and fades out for 1
second
*5: Fades in for 2 seconds, and then turns off
Содержание CXDI-55C
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...