2. Installation
15
3)
Connect the connector of the remote cable to the power supply and joint the skin top spiral
and lock nut temporarily. Adjust the length of the cable with some margin in length for
wiring. After adjusting the length properly, fasten the skin top spiral tightly.
Note: To avoid the risk of damage when the cables are removed with very large force,
check skin top spiral of the bush (refer the following figure).
<Remote cable>
<X-RAY I/F cable>
4)
Fasten the skin top spiral and lock nut tightly using 15mm screw wrench.
Gap 1.0 to 1.5mm
No gap
Skin top spiral
Lock nut
Connector