77
TE-BBR Instruction and Operation Manual
Appendix 3 – Technical Note 0611091
BBR Crack Seal Option Installation
This technical note describes the procedure for installing the Crack Seal Option (p/n 44.0675) for the
Thermoelectric Bending-Beam Rheometer. Refer to Table 9 for components of the Crack Seal kit.
Table 9: BBR Crack Seal Kit Components (p/n 44.0675)
Description
Part Number
Quantity
Binder with Calibration Certificates
93.2021
1
6-32 ½-inch Phillips flat-head screws
01.0920
2
6-32 3/8-inch Phillips flat-head
screws
01.0880
2
Support, Crack Seal
44.0659
2
Adaptor, Support, Crack Seal
44.0660
2
Thin beam assembly for Crack Seal
verification
44.0663
1
New ½” thick beam for Crack Seal
calibration
44.0664
1
Application Note 0611091-Crack Seal
Installation Instructions
44.0910
1
Tools Required:
Stubby #2 Phillips screwdriver, stubby flat blade screwdriver, absorbent paper
towels
Replacement Procedure
Caution:
Turn off TE-BBR and remove MAINS power from the instrument heat
exchanger power supply before beginning procedure. Instrument is heavy, use
two or more people for this procedure.
1.
Disconnect the Load Frame from the Control Unit. Remove the four bayonet signal connectors and
the two air line connections at the rear of the Load Frame.
Содержание TE-BBR
Страница 1: ...Instruction Operation Manual 44 0991 Thermo Electric Bending Beam Rheometer TE BBR ...
Страница 2: ......
Страница 10: ......
Страница 11: ......
Страница 44: ...CANNON Instrument Company Software Menu Options 33 Figure 17 Toolbar Options ...
Страница 47: ...36 TE BBR Instruction and Operation Manual Figure 18 Mold Specimen Diagram ...
Страница 70: ...CANNON Instrument Company Troubleshooting 59 Figure 25 Diagram A new Figure 26 Diagram B old ...
Страница 85: ...74 TE BBR Instruction and Operation Manual Figure 31 Graph View ...