28
TE-BBR Instruction and Operation Manual
Verify Calibration
To verify that the instrument is providing accurate data for force and deflection complete the
procedure outlined in this section.
1.
With the instrument in Local
and Zero
modes, place the gage block in the load frame with the
shaft on the top level of the block.
2.
Read the deflection from the computer screen and verify that the value agrees with the
certificate of calibration for the gage block (
∆
RED) ± TOL.
3.
Repeat the previous two steps for the red, white and blue levels.
4.
Place the shaft on the red level of the gage block and read the force. The value should be
0.00 ± 10 mN.
5.
Remove the gage block from the Load Frame.
6.
Adjust the Zero regulator until the shaft is floating freely at the approximate midpoint
in its vertical travel.
7.
Gently add a weight of 2 grams to the load platform
and
verify that the shaft drops slowly
downward under the weight. Refer to
Figure 16: Placing Weight with Forceps
8.
Remove the weight and place the quarter inch beam on the beam supports in the bath.
9.
Verify that the instrument is in the Zero
mode and adjust the Zero regulator until the shaft is
touching the beam with a force of 20 mN ± 10 mN.
10.
Add one 2-gram weight (± 0.2 g) to the load platform.
11.
Verify that the force increases by 20 mN ± TOL.
12.
Add a second 2-gram weight (± 0.2 g) to the load platform.
Содержание TE-BBR
Страница 1: ...Instruction Operation Manual 44 0991 Thermo Electric Bending Beam Rheometer TE BBR ...
Страница 2: ......
Страница 10: ......
Страница 11: ......
Страница 44: ...CANNON Instrument Company Software Menu Options 33 Figure 17 Toolbar Options ...
Страница 47: ...36 TE BBR Instruction and Operation Manual Figure 18 Mold Specimen Diagram ...
Страница 70: ...CANNON Instrument Company Troubleshooting 59 Figure 25 Diagram A new Figure 26 Diagram B old ...
Страница 85: ...74 TE BBR Instruction and Operation Manual Figure 31 Graph View ...