Calibration
To provide accurate readings from the thermistor probe, it is necessary to calibrate the probe
periodically with a calibrated thermometer.
Thermistor Probe Calibration Procedure
1.
Load the BBRw software and select
Calibrate
> Thermistor
from the BBRw Main Menu or use
the Thermometer icon.
2.
Enter the current temperature of the bath to the nearest 0.01 °C as indicated by a separate
calibrated thermometer and click
OK
.
LVDT and Load Cell Calibration
It is vital to calibrate the LVDT (Linear Variable Differential Transducer) and Load Cell in order to
expect accurate readings for the output.
1.
Select
Calibrate
from the BBRw Main Menu.
2.
Select
Force
,
Deflection
and
Compliance
from the calibration options.
3.
Follow the on-screen prompts to complete calibration.
A cylindrical metal gage block is provided for the LVDT calibration. Refer to Figure 17. This block is
placed to the left of the load platform on the small pin provided so the block can be rotated easily. A
6.35 mm (¼") thick stainless steel beam and four 100 gram weights are supplied for the calibration.
At the end of the calibration run, remove the weights from the load platform, and the 6.35 mm (¼")
steel beam from the bath.
Содержание TE-BBR
Страница 1: ...Instruction Operation Manual 44 0991 Thermo Electric Bending Beam Rheometer TE BBR ...
Страница 2: ......
Страница 10: ......
Страница 11: ......
Страница 44: ...CANNON Instrument Company Software Menu Options 33 Figure 17 Toolbar Options ...
Страница 47: ...36 TE BBR Instruction and Operation Manual Figure 18 Mold Specimen Diagram ...
Страница 70: ...CANNON Instrument Company Troubleshooting 59 Figure 25 Diagram A new Figure 26 Diagram B old ...
Страница 85: ...74 TE BBR Instruction and Operation Manual Figure 31 Graph View ...