3.
Watch the Force value on the
Instrument Status Bar
and adjust the Control Unit ZERO
pressure dial until the load reading displayed is 35 mN ± 10 mN.
4.
Switch to the LOAD
mode (Control Unit front panel, left side) and adjust the LOAD
regulator
until the computer display reading for the load is 980 ± 50 mN.
5.
Switch back to the ZERO
mode and check to see if the shaft has drifted up off the beam. If so,
repeat the load procedure.
6.
Manually raise the shaft enough to remove the stainless steel beam.
7.
Manually raise the haft enough to remove the stainless steel beam.