CANNON Instrument Company | Initial Software Settings
23
Initial Software Settings
Before running TE-BBR tests, setup information must be entered.
1.
Turn on computer and load the BBRw software.
2.
Select
Instrument Setup
from the BBRw menu options. Refer to Figure 16.
Figure 14: Instrument Setup
3.
Set the COM port:
a.
Click on
Port
from the tab options on the
Instrument Setup
menu.
b.
Click on (down arrow) from the Serial Port page to display the port selection options.
c.
Click on the Serial Port setting which corresponds with the port for the RS-232 cable
connection between the BBR and your PC.
4.
Input equipment data, including thermistor calibration constants:
a.
Click on
Equipment
from the tab options on the
Instrument Setup
property sheet.
b.
Input an alphanumeric identifier for your BBR instrument (up to 30 characters in length) in
the
Apparatus ID
field.
c.
Input the thermistor probe values (constants) for
S/N
,
A0
,
A1
and
A2
from the calibration
certificate shipped with the probe. Use the tab
key or mouse to select the appropriate fields
for data entry. Make sure to include negative signs (-) and decimal points as indicated on the
calibration certificate.
Note:
Once set, the thermistor constants should only be changed if replacing the
thermistor.
Содержание TE-BBR
Страница 1: ...Instruction Operation Manual 44 0991 Thermo Electric Bending Beam Rheometer TE BBR ...
Страница 2: ......
Страница 10: ......
Страница 11: ......
Страница 44: ...CANNON Instrument Company Software Menu Options 33 Figure 17 Toolbar Options ...
Страница 47: ...36 TE BBR Instruction and Operation Manual Figure 18 Mold Specimen Diagram ...
Страница 70: ...CANNON Instrument Company Troubleshooting 59 Figure 25 Diagram A new Figure 26 Diagram B old ...
Страница 85: ...74 TE BBR Instruction and Operation Manual Figure 31 Graph View ...