o
-CELL215 minimum = 20 mA, average = 75 mA
o
-CELL220 minimum = 20 mA, average = 75 mA
o
-CELL225 minimum = 20 mA, average = 75 mA
Wi-Fi Additional Current Contribution at 12 VDC:
l
Client mode communicating: 70 mA typical
l
Client mode idle: 7 mA typical
l
Access point mode communicating: 70 mA
l
Access point mode idle: 62 mA typical
l
Idle (use
IPNetPower()
or DevConfig setting to disable): <0.1 mA
RF Average Additional Current Contribution at 12 VDC
-RF407, -RF412, -RF427
-RF422
Transmit
< 80 mA
20 mA
Idle On
12 mA
9.5 mA
Idle 0.5 s Power Mode
4 mA
3.5 mA
Idle 1 s Power Mode
3 mA
2 mA
Idle 4 s Power Mode
1.5 mA
1.5 mA
See also
19.4 Power output specifications
System power out limits (when powered with 12 VDC). Current is limited by a self-resetting
thermal fuse.
l
5.8 A @ -40 °C
l
3.7 A @ 20 °C
l
2.0 A @ 70 °C
VX: Two independently configurable voltage terminals (VX1-VX2). VX outputs are produced by a
12-bit DAC. In this case, these terminals are regularly used with resistive-bridge measurements
(see
(p. 72) for more information). VX terminals can also be used to
supply a switched, regulated 5 VDC power source to power digital sensors and toggle control
lines.
19. CR300 series Specifications
186