OvC
:
the channel is in Over Current condition, i.e. the channel has
reached the current limit.
•
UnV
:
the channel is in Under Voltage condition, i.e. the actual value
of the channel output voltage is lower than the programmed
value.
•
OvV
:
the channel is in Over Voltage condition, i.e. the actual value
of the channel output voltage is higher than the programmed
value.
•
HVmax
:
the channel output voltage has reached the HVMAX hardware
voltage output limit.
This field becomes red if the channel is not in a steady status (normal operation). If the
system is operated from an external terminal it blinks instead of being red.
7.1.9.2 Board Status
The possible board status which can be displayed in the
Bd Status
column of the
Channels Window
are as follows:
•
Pw Fault
:
power supply fault on the board.
•
Fw Err
:
firmware error on the board.
•
VMAX Cal Err
:
calibration error for the HVMAX parameter of the board.
•
T Cal Err
:
calibration error for the temperature detection on the board.
This field becomes red if the board is not in a steady status (normal operation). If the
system is operated from an external terminal it blinks instead of being red.