Performance checking
MXT Transmitter
63
12. Set the Transmitter to full power, ‘Continuous’
output and ‘33kHz’.
13. Press the ‘M’ button on the Function Checker to select
‘Con’ (connected).
14. Plug the Direct Connection Leads (red and black) into
the socket on the Transmitter and check the Connected
Mode icon is displayed at the top of the display. Attach a
crocodile clip to each terminal on the Function Checker.
CORRECT RESULT
: The audio response should change to
a lower pitch as the second clip is attached to the second
terminal and the Transmitter bar graph should indicate a
full scale response. The Function Checker bar graph should
show an indication of 8/10 full scale and the ‘33kHz’ icon
should show in the top left hand corner on the Function
Checker display.
NOTE
If the CORRECT RESULT is not achieved, the cause may be a set of faulty Direct Connection Leads.
Replace with good leads and try the test again.