55
14. Plug the yellow 10m Auxiliary Earth Lead into
the green socket on the Signal Generator and attach
the crocodile clip to the unused terminal on the
Function Checker.
CORRECT RESULT
: The audio response should
change to a lower pitch as the second crocodile
clip is attached to the second terminal.
The Signal Generator bar graph should indicate
a full scale response. The Function Checker bar
graph should show an indication of 5/10 full
scale (for the black SGV) or 8/10 full scale (for
the grey 1 Watt SGV) and the ‘33kHz’ icon should
show in the top left hand corner on the Function
Checker display.
NOTE
If the CORRECT RESULT is not achieved, the cause may be a faulty 10m Auxiliary Earth Lead.
Replace with a good lead and try the test again.
If all of the ‘Condition Checks’ (pages 11-12) and ‘Performance Checks’ (pages 50-55) have been completed
successfully then the Signal Generator is operating correctly.
SGV Signal Generator
Function Checker
Function Checker
Performance checking
SGV Signal Generator
- black/grey
Function Checker display
when checking BLACK SGV
Function Checker display
when checking GREY SGV