Performance checking
MXL
32
8. Set the Sensitivity Control fully anti-clockwise and slide
the Locator along the side of the Function Checker to find
the position of maximum signal.
CORRECT RESULT
: The Locator bar graph should show
a mid-scale response with the audio tone sounding.
NOTE
It is important to NOT move the Locator from this position for the remaining Performance Checks.
NOTE
If the bar graph does not reduce to a mid-scale response when on minimum sensitivity it may be due to
background electromagnetic signals. Refer to the ‘Selecting a suitable location’ section on page 8.
9. Check the correct operation of the Sensitivity Control
by adjusting it first fully clockwise (maximum sensitivity)
and then fully anti-clockwise (minimum sensitivity).
CORRECT RESULT
: The Locator bar graph should change from
showing a full-scale response when on maximum sensitivity
to a mid-scale response when on minimum sensitivity.
NOTE
It is important to NOT adjust the sensitivity setting of the Locator for the remaining Performance Checks.
<
<
MXL Locator
MXL Locator
MXL Locator
<
<
3/4
turn
3/4
turn
<
3/4
turn