36
GasAlertQuattro
Troubleshooting
Startup Troubleshooting
Error Screen
Problem
Solution
Error Screen
Problem
Solution
Auto-zero Error
Sensor(s) failed to
auto-zero
Calibrate the sensor.
Last Calibration
Failed
Displays when the
last calibration failed.
If the
Force
Calibration
option is
enabled, the sensors
must be calibrated.
Press
C
and calibrate the
sensor(s) immediately.
Refer to
Calibration on
page 10
.
If
Cal
IR Lock
is enabled,
an IR device (IR Link or
MicroDock II) must be
used to calibrate.
Self-test Failed
Sensor(s) fails the
self-test during
startup.
Press
C
to accept the
failed sensor(s). The
Sensor self test error
accepted
screen
displays. Replace the
sensor when startup is
complete. Refer to
Replacing the Sensors on
page 24
.
Force Calibration
If the
Force
Calibration
option is
enabled, the sensors
must be calibrated to
enter normal
operation.
Press and hold
C
to
calibrate the sensors, or
press
C
and release to
deactivate the detector.
Refer to
Calibration on
page 10
.
If
Cal IR Lock
is enabled,
an IR device (IR Link or
MicroDock II) must be
used to calibrate.
All Sensors Fail
If the
Lockout on
Self-test Error
option
is enabled and all
sensors fail, the
detector
automatically
deactivates.
A failure of all sensors
can be caused by
poisoning (alcohol and
silicon). Allow sensors to
recover for 1 hour. If
sensors fail startup again,
refer to
Replacing the
Sensors on page 24
.
Calibration Overdue
Displays when
calibration is past
due. If the
Force
Calibration
option is
enabled, the
sensor(s) must be
calibrated to enter
normal operation.
Press
C
to continue and
calibrate the sensor(s)
immediately. Refer to
Calibration on page 10
.
If
Cal IR Lock
is enabled,
an IR device (IR Link or
MicroDock II) must be
used to calibrate.
H
2
S
CO
LEL
O
2
Zeroing ERROR
H
2
S
CO
LEL
O
2
Last Calibration
FAILED
waiting... 272
Press to accept
H
2
S
CO
LEL
O
2
Sensor self test
ERROR
Press to accept
waiting... 272
Force Calibration
enabled...
Hold to off
or continue to
hold to Cal.
waiting... 272
All sensors
FAILED self test
Powering off
H
2
S days
CO days
LEL
O
2
days
180 180
180
CAL
DUE
Press to skip
waiting... 272
Содержание GasAlertQuattro 1
Страница 1: ...Quick Reference Guide 1 2 3 and 4 Gas Detector ...
Страница 4: ......
Страница 11: ...7 GasAlertQuattro Screen Elements Screen Elements ...
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ...50104941 536 EN Rev F 2017 BW Technologies by Honeywell All rights reserved ...