38
GasAlertQuattro
Quick Reference Guide
Calibration Troubleshooting
Bump Test Troubleshooting
For bump test troubleshooting recommendations, refer to the Startup Troubleshooting table.
Error Screen
Problem
Solution
Error Screen
Problem
Solution
Detector
deactivates while
trying to calibrate
Detector does not
enter calibration.
Displays OFF and
then deactivates.
Calibrate the sensor.
Cal IR Lock Enabled
IR lock enabled
screen displays.
An IR device must be used
to calibrate (IR Link or
MicroDock II). For manual
calibration, refer to
Calibrating Using an IR
Device
in the
GasAlertQuattro User
Manual
. For automated
calibration, refer to the
MicroDock II User
Manual
.
Self-test Fail
If a sensor(s) fails
auto zero, an error
message displays
showing which
sensor(s) failed.
Replace the sensor or
contact BW Technologies
by Honeywell. Refer to
Replacing the Sensors on
page 24
Calibration Error
Insufficient amount of
gas detected.
Verify the span gas values
on the cylinder match the
span gas values set for the
detector. Ensure gas is
applied at a flow rate of
250-500 ml/min. Ensure
cylinder is not empty or
expired. Replace
immediately if required.
Replace the regulator if
required.
OFF
Calibration
IR lock enabled...
Connect to
MicroDock or
Fleet Manager
Press to end
waiting... 24
H
2
S ppm
CO ppm
O
2
%
0
0.0
20.9
LEL %
Calibration
gas not
detected
Press to skip
Содержание GasAlertQuattro 1
Страница 1: ...Quick Reference Guide 1 2 3 and 4 Gas Detector ...
Страница 4: ......
Страница 11: ...7 GasAlertQuattro Screen Elements Screen Elements ...
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ...50104941 536 EN Rev F 2017 BW Technologies by Honeywell All rights reserved ...