Specifications
Instrument dimensions:
13 x 8.1 x 4.7 cm
(5.1 x 3.2 x 1.5 in.)
Weight:
316 g (11.15 oz.) with rechargeable battery pack
338 g (11.92 oz.) with alkaline battery pack
Operating temperature:
-20°C to +50°C (-4°F to +122°F)
Storage temperature:
-40°C to +60°C (-40°F to +140°F)
Operating humidity:
10% to 100% relative humidity
(non-condensing)
Operating environment specifications for use according to
European performance approvals
(Oxygen and methane measurements)
Operating temperature range certified by BAM:
-20°C to
+50°C
Operating humidity as tested by BAM:
5% r.H. to 95% r.H.
(Extended Range of Use for Temperature and Humidity as
Compared To EN 50104 (Oxygen performance) and
EN 67009-29-1 (LEL performance))
Storage temperature range tested by BAM:
-25°C to +60°C
Operating pressure as tested by BAM:
80 kPA to 120 kPa
Storage duration:
Two years from purchase date
Dust and moisture ingress:
IP66/67 (with screw engaged)
Alarm setpoints:
May vary by region and are user-defined
Detection range:
H
2
S: 0 - 200 ppm (0.1 ppm increments from 0.0 to 39.9 ppm/
1 ppm increments above 40 ppm)
CO: 0 - 1000 ppm (1 ppm increments)
O
2
: 0 - 30.0% vol. (0.1% vol. increments)
Combustible (LEL): 0 - 100% (1% LEL increments) or
0 - 5.0% v/v methane
Sensor type:
H
2
S, CO, O
2
: Single plug-in electrochemical cell
Combustibles: Plug-in catalytic bead
O
2
measuring principle:
Capillary controlled concentration
sensor
Bump test specified limits:
BW recommends using a gas
cylinder that will ensure the combustible sensor has an
accuracy of -0 to +20% of actual reading (reference CAN/CSA
C22.2 No. 152)
Alarm conditions:
TWA alarm, STEL alarm, low alarm, high
alarm, multi alarm, over limit (OL) alarm, low battery alarm,
critical low battery alarm, sensor failure alarm, IntelliFlash,
confidence/compliance beep
Audible alarm:
95 dB at 30 cm (12 in.) variable pulsed beeper
Visual alarm:
Red light-emitting diodes (LEDs)
IntelliFlash:
Green light-emitting diode. Flash frequency is
user-defined with IntelliFlash interval option
Confidence/compliance beep:
Audible beep from variable
pulsed beeper. Beep frequency is user-defined with confidence/
compliance beep interval option
Minimum performance compliance:
IntelliFlash should be set
to no slower than 4 seconds to comply with European
Regulations
Содержание GasAlertQuattro 1
Страница 1: ...Quick Reference Guide 1 2 3 and 4 Gas Detector ...
Страница 4: ......
Страница 11: ...7 GasAlertQuattro Screen Elements Screen Elements ...
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ...50104941 536 EN Rev F 2017 BW Technologies by Honeywell All rights reserved ...