Instruction Manual for Vertica-Homecare Bed
Page
34
4.3.6
Activating Stand-Up Function
1
+
Plug in mains cable + brake castors
(Stand-up function only works with
mains power operation and with
braked castors for safety reasons).
2
Activate marked buttons on the NURSE
handset
[15]
at the same time for at
least 3 seconds
):
NURSE handset [15] is released
Patient handset [3] and control
panel [13] are locked
Hold down the buttons again for at
least 3 seconds to relock this function.
3
Stand-up function display active
by:
Orange blinking light on NURSE
handset
When the stand-up function is active, the patient handset is
completely locked for safety reasons.
When the stand-up function is active on the NURSE handset [15], the
CPR emergency function
can also be adjusted.
The stand-up function remains active until the bed has returned to the
horizontal starting position. See Chapter 4.3.9