![burmeier Vertica-Homecare Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/burmeier/vertica-homecare/vertica-homecare_instruction-manual_2832796064.webp)
Instruction Manual for Vertica-Homecare Bed
Page
65
6.5.6
RESET Control Unit CB16
When should it be carried out?
After installing / exchanging the control unit CB16
To unlock the control unit CB16 after functional failures.
How should this be done?
On the patient handset:
Press down on both of the marked keys “Backrest
UP+DOWN”
simultaneously for 5 seconds until the
“Beep” signal goes off
Effect:
Deletes any possible errors in CB16 (reset)
Release of the locked control
Afterwards, absolutely also perform: Actuator alignment; see Chapter
6.5.7
Actuator Alignment
During the alignment, the current start positions of the actuators are transmitted
to the control unit through a reference run to “0” and then the existing tolerances
in ongoing operation are aligned after a period.
Valid for
: all actuators M1 to M4 (not necessary for actuator M5: backrest)
When should this be carried out?
After exchanging the control unit, the actuator or after a RESET.
If the module emits regular sound signals when making electrical
adjustments as well as when connecting the bed to the mains power,
and/or no adjustments are possible, or adjustments are only possible on
one side.
When carrying out an alignment, it is important that none of the
functions are locked on the patient handset!
Any bed positions which have been saved remain saved.
Содержание Vertica-Homecare
Страница 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Страница 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Страница 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Страница 83: ...Instruction Manual for Vertica Homecare Bed Page 84...