Instruction Manual for Vertica-Homecare Bed
Page
20
4.2
E
LECTRIC
D
RIVE
S
YSTEM
4.2.1
Special Safety Information on the Electrical Actuator System
This hospital bed may not be used in combination with high-
frequency surgical equipment or in explosive environments!
When using accessories:
Make sure that the attachment of accessories does not produce
any crush or shearing zones when the bed sections are adjusted. If
this cannot be ensured, the user must lock the handset using the
control panel or the locking box!
When making any adjustments, always ensure that there are no
limbs belonging to patients, users, other persons, and especially
playing children, that could be trapped underneath the rests or
the mattress base.
Ensure that the mains cable and handset cable cannot be driven
over or otherwise crushed when the bed is moved.
Always keep the handset in a safe place, where the housing and
buttons cannot be damaged mechanically, particularly as a
result of collision with other objects. Hang the handset on the hook
or holder provided, within the protected bed area. This helps to
protect patients from hazards arising from functional defects and
unintentional movements.
Ensure that no obstacles, such as bedside lockers, supply rails, wall
protection rails, window sills etc. or chairs could impede
adjustments to the bed.
Ensure that there are no objects on the chassis.
Otherwise, mechanical damages could occur to the bed and/or
related objects which could impede safety.
When routing the handset cable, ensure that it cannot be
damaged by any moving parts of the bed.
Ensure that the mains cable and handset cable cannot be driven
over or otherwise crushed when the bed is moved.
Continuous operation must not exceed two minutes! After this
time, a rest period of at least 18 minutes must be observed.
Alternatively: one minute continuous operation followed by a nine
minute rest period.
If the electric drive is operated for a much longer period, e.g. due
to the patient "playing" continually with the handset, the thermal
protection device integrated in the transformer of the control unit
will deactivate the drive permanently. In this extremely rare case,
the control unit must be replaced.
Содержание Vertica-Homecare
Страница 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Страница 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Страница 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Страница 83: ...Instruction Manual for Vertica Homecare Bed Page 84...