Instruction Manual for Vertica-Homecare Bed
Page
24
4.2.2.2
Locking Function
Only users are authorised to operate the locking function! The medical
personnel must decide which functions, if any, should be locked.
If the clinical condition of the patient is so critical that adjustments
using the handset might be dangerous for her/him, the user must
immediately lock the respective function(s) using the control
panel.
The bed remains in the position that it was in at the moment when
the functions were switched off.
A sticker on the side panel on the long side of
the bed indicates:
Locking is especially necessary if, for example,
-
If the patient is not able to operate the bed safely or to
extricate himself from dangerous situations.
-
If the patient could be harmed by unintentional or unexpected
adjustment of the drives.
-
Accessories or other devices are attached that could restrict
the adjustment range of the bed, and put the patient at risk or
cause damage to equipment.
-
The patient is exposed to an increased risk of entrapment during
backrest and thigh rest adjustments when the safety sides are
raised,
-
Children are left unattended in a room with the bed.
In these cases, adjustments must only be performed by, or in the
presence of a trained person!
-
The user must check to ensure that the handset is locked/enabled by pressing
the corresponding buttons after:
o
Re-enabling or re-locking.
o
Before using a bed again after it has been out of operation.
o
Before using a bed after machine washing or disinfection has taken place.
-
The locking keys are on the handset cable when the bed is
delivered (transport securing device). Before operating the bed,
the locking keys are to be removed from the cable and put out
of the range of patients or other parties (such as key ring users /
carers). This avoids unauthorised actuating of the locking device
and consequent, potential patient endangerment.
-
Do not forcibly turn the locking key beyond the stop of the lock-
ing surface! The lock or the handset can be damaged.
Содержание Vertica-Homecare
Страница 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Страница 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Страница 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Страница 83: ...Instruction Manual for Vertica Homecare Bed Page 84...