Instruction Manual for Vertica-Homecare Bed
Page
57
6.2
R
EGULAR
S
AFETY
I
NSPECTIONS ON
C
ARE
B
EDS
The operator of this care bed is obliged according to MPBetreibV (Medical
Devices Operator Ordinance) Section 4 to conduct regular inspections in each
new building, after all maintenance work and during operation to ensure the safe
condition of the care bed.
These inspection are to be repeated within the regular maintenance activities
depending on the conditions of use according to MPBetreibV Section 4 and the
inspections prescribed by the Employers' Liability Insurance Associations for mobile
electrical equipment in commercial use according to DGUV A3 (Testing of Mobile
Electrical Equipment in Commercial Use).
Inspection Cycle
We recommend, as a guideline, that an annual DGUV A3 inspection be carried
out by our qualified service engineers, with verification of adherence to the 2%
error rate (see also the DGUV A3: § 5, Table 1B).
Observe this order for the inspection according to EN 62353 (VDE 0751):
I.
Visual inspection
II.
Electrical measurement
III.
Performance inspection
Visual and functional check
-
The visual inspection and function testing as well as the assessment and docu-
mentation of the test results must be conducted exclusively by competent
persons, according to MPBetreibV Section 4 and DIN EN 62353 (VDE 0751), who
have the required qualifications and tools for proper inspections and testing.
Electrical measurement
-
The electrical measurement according to VDE 0751 may also be conducted by
electrically instructed persons (in the sense of DGUV A3) with additional
medical and device-specific if appropriate measuring instruments are present.
-
The assessment and documentation of the test results may only be made by a
qualified electrician with medical and device specific additional knowledge.
The protocol template for inspections based on the EN 62353 (VDE 0751) standard
is found on the following pages.
If damage or malfunctions are found, the bed must be taken out of
operation immediately until it has been repaired or the defective
components have been replaced!
A repeat inspection must be carried out in accordance with the test
sheet to determine whether the damage or malfunction has been
rectified.
Only then have the requirements for further operation been met.
Содержание Vertica-Homecare
Страница 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Страница 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Страница 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Страница 83: ...Instruction Manual for Vertica Homecare Bed Page 84...