![burmeier Vertica-Homecare Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/burmeier/vertica-homecare/vertica-homecare_instruction-manual_2832796074.webp)
Instruction Manual for Vertica-Homecare Bed
Page
75
9.4
A
MBIENT
C
ONDITIONS
The following ambient conditions must be maintained:
For storage / transport:
Minimum
Maximum
Temperature:
-10° C
+ 50° C
Relative humidity:
20 %
90 %
At 30°C; non-condensing
Air pressure:
800 hPa
1060 hPa
At altitude ≤ 2000m
In operation:
Minimum
Maximum
Temperature:
+ 5° C
+ 40° C
Relative humidity:
20 %
90 %
At 30°C; non-condensing
Air pressure:
800 hPa
1060 hPa
At altitude ≤ 2000m
9.5
T
ECHNICAL
I
NFORMATION ON
E
LECTROMAGNETIC
C
OMPATIBILITY
(EMC)
To ensure EMC, only use cables and accessories approved by the manufacturer
(see Chapter 9.3)
The use of accessories, sensors or cables other than those approved, with the
exception of sensors and cables sold by the equipment manufacturer as
replacement parts for internal components, can result in an increase in the
transmission level or a reduction in the immunity level of the equipment.
The equipment may not be used directly next to or on top of other equipment.
If it is necessary to use the equipment in this way, you must check to ensure
that it functions properly in the required configuration.
Guidelines and Manufacturer’s Declaration – Electromagnetic Emissions –
The BED is intended for use in the electromagnetic environment described below. The operator
or user of the BED should ensure that it is used in such an environment.
Interference emission
measurements
Compliance
Electromagnetic environment - guidelines
HF emissions
to CISPR 11
Group 1
The BED uses HF energy for its internal functions
only.
HF emissions
to CISPR 11
Class B
The BED is intended for use in all types of
establishment including residential and the like
that are directly connected to a public supply
network that also serves buildings that are used
for residential purposes.
Harmonics
according to IEC 61000-3-2
Class D
Voltage fluctuations/
flicker acc. to IEC 61000-3-3
Complies
HF emissions
according to CISPR 14-1
Complies
The BED is not intended for connection to other
technical equipment.
Содержание Vertica-Homecare
Страница 1: ...Instruction Manual Care bed with stand up function Vertica Homecare Date 2018 05 25 241335 v3...
Страница 81: ...Instruction Manual for Vertica Homecare Bed Page 82 Notes...
Страница 82: ...Instruction Manual for Vertica Homecare Bed Page 83 Notes...
Страница 83: ...Instruction Manual for Vertica Homecare Bed Page 84...