Trouble-shooting table aid
Problem
Possible causes
Solution
hand switch / drive
system not functioning
n
hand switch plug not properly
connected
n
Transformer plug (24 volt) not
inserted properly
n
hand switch or drive system
defective
n
functions are blocked on
hand switch
F
Test the plug connection
F
Press the transformer unit
plug firmly into the connecting
socket
F
inform your operator about
necessary repairs
F
Unlock functions (see
page 23)
green Led on trans-
former does not light up
and drive systems do not
function
n
Power cable not properly
connected
n
Wall socket without power
n
Transformer defective
F
Plug in the power cable
properly
F
Test the wall socket and fuse
box
F
exchange the transformer;
inform your operator about any
necessary repairs
green Led on trans-
former lights up continu-
ously, yellow Led does
not light up when any
hand switch button is
pressed
n
hand switch is defective
n
Transformer plug (24 volt) not
inserted properly
n
functions are locked on hand
switch
F
exchange the hand switch
F
Test the plug connection
F
Unlock the functions
(see page 23)
drives operate in one
direction only
n
hand switch or drive system is
defective
F
inform your operator about
necessary repairs
drives stop suddenly
after prolonged adjust-
ment times
n
heat swich in transformer is
activated
F
Pay attention minimum pause
(see page 22)
F
exchange the transformer;
inform your operator about any
necessary repairs
Mattress base not
horizontal
n
Lifting actuators not running
synchronously
F
raise or lower mattress base
to highest or lowest setting at
least once a day.