87
English
SE58
L
[022] BATTERY V>MAX
Battery voltage (rechargeable)>
max. Allowed
Contact the service to Replace the battery
[023] BATTERY I>MAX
Battery charge current> max.
allowed
Contact the service to Replace the battery
[024] MAIN PS V.ERR
Main supply voltage (+ 5V) out of
tolerance.
Contact the service
[025] USB VOLTAGE ERR Voltage of USB connection out of
tolerance.
Contact the service
[026] SDC ALMOST
FULL
SD card space <500 MB.
For more information see function “12.9”
[027] SDC FULL
SD card out of memory
Memory Full. You can not save logger.
Contact the service to replace the SD
memory.
[028] BATT.TEMP.CRIT
The battery can not be charged.
The temperature is out of range
( detected temperature <0 C° or
temperature >50°)
Wait for the normal temperature reset.
View Environmental Use Conditions
“Environmental Use Conditions” pag. 5.
[030] BATCHING ERROR
This error activates in two
conditions
(see “BATCHING” page 52)
Check the status of the actuator and the
system