5
English
SE58
L
Power supply
versions
Power supply
voltage
Power supply
frequency
Min
Power
Max power
HV
100-240V~
45-66HZ
1,5 W
(Sensor only)
12 W
(all Loads)
LV
24-36V
//
24-36V~
45-66HZ
LLV
12-48V
//
The transmitter can be installed internally or externally
Altitude: untill 4000m
Humidity range: 0-100%
If the transmitter is supplied in compact version (transmitter over the sensor), consider the ambient temperatures
more restrictive, otherwise refer to the relevant manuals.
* For discontinuous use, a thermostat heat source installation may be necessary.
TECHNICAL CHARACTERISTICS
Electrical Characteristic
Environmental Use Conditions
ATTENTION
The battery will not be charged outside the below limits:
T board SE58 M < 0 °C
T board SE58 M > 50 °C
AMBIENT TEMPERATURE
Aluminium
Reinf. Nylon
Min*
Max
Min*
Max
°C
-20
+ 60
-10
+ 50
°F
-4
+140
+14
+122
Converter classification: class I,
IP67 (where: 6 = totally protected against dust, 7 = Protected against the effects of temporary
immersion) /
IP68 (where: 6 = totally protected against dust, 8 = continuous immersion 1.5 m; 1 h)
for aluminum and PA6 housing, installation category (overvoltage) II, rated pollution degree 2.
• Voltage variations must not exceed ±10% of the nominal one.
• Input/output insulated
• The output 4-20mA (optional) is electrically connected to the ON/OFF outputs and the output power supply
(24V
).
• Version LV/LLV : inrush current < 20A
Version HV : inrush current < 25A
Note:
The devices powered at 24-36 V and 12-48 V must only be powered with power supplies compliant with the
IEC61010 standard
Содержание B0A1C2B1A4A0
Страница 1: ...Type SE58 L Transmitter Operating Instructions...
Страница 8: ...7 English SE58 L Dimensions 200 114 164 167 114 180 290 195...
Страница 9: ...8 English SE58 L Remote version 142 28 295 120 7 30...
Страница 11: ...10 English SE58 L DIMENSIONS STAINLESS STEEL VERSION Compact Version 171 156 186 146...
Страница 12: ...11 English SE58 L Separate Version 236 40 7 90 241 55...