When a measure mass unit is set, the specific gravity function is automatically enabled by the system. Please,
note that the mass measure is heavily affected by the temperature. With certain liquids this may cause significant
measurement errors. The following measure of time units can be selected: s = second, m = minute, h = hour,
d = day .
MASS UNIT
METRIC
g
Gram
kg
Kilogram
t
Ton
MENU 3 - SCALE
The FS2 (full scale flow rate 2) is active with POS. 6.6 see page 35, enabled.
The full scale is used to indicate the maximum meter’s flow rate. The full scale should be chosen carefully as its
parameters are used for several other parameters.There are three fields to fill in order to set this parameter, from
left to right: 1) measure unit, 2) time unit of measure and 3) numeric value. The selection is made by positioning
the cursor on the field to modify. To change the unit of measure type (metric, Imperial units, mass or volume) see
POS. 2.3 see page 34 and POS. 2.19 see page 34 and POS. 2.20 see page 34. The value of full scale 1
and 2 depends on nominal diameter POS. 1.4 see page 34. The following tables show the units of measure
available and the conversion factor by comparison with 1dm3 and 1kg. The transmitter accepts any kind of
combination of units of measure satisfying both the following conditions:
Numeric field values 99999
1/25 fsmax ≤ numeric field value ≤ fsmax.
Where fsmax is the maximum full scale value corresponding to the sensor, equal to a 10m/s liquid speed. The
measure units are shown as appear on the display. The imperial units units are diversified by using capital and
small characters.
NOT METRIC
in³
Cubic inch
Gal
American gallon
ft³
Cubic foot
bbl
Standard barrel
BBL
Oil barrel
hf³
Hecto cubic feet
KGL
Amer
i
can Kilo gallon
kf³
Kilo cubic feet
ttG
Ten thousand gallons
IGL
Imperial gallon
IKG
Imperial Kilo gallon
Aft
Acre foot
MGL
Mega gallon
IMG
Imperial mega gallon
METRIC
cm³
Cubic centimeter
ml
Milliliter
l
Liter
dm³
Cubic decimeter
dal
Decalitre
hl
Hectolitre
m³
Cubic meter
ML
Mega Liter
MASS UNIT NOT
METRIC
Oz
Ounce
Lb
Pound
Ton
Short tons
(POS. 3.1-3)
Flow Rate Full Scale 1-2
[FS1-2= l/s xxxx.x]
AL2
[FRFS1-FRFS2]
The command FRFS1 =? and command FRS2 = ?, edited by MCP software, return a list of only the unit
compatible with the nominal diameter set.
NOTES FOR USING THE MCP INTERFACE
Содержание B0A1C2B1A4A0
Страница 1: ...Type SE58 L Transmitter Operating Instructions...
Страница 8: ...7 English SE58 L Dimensions 200 114 164 167 114 180 290 195...
Страница 9: ...8 English SE58 L Remote version 142 28 295 120 7 30...
Страница 11: ...10 English SE58 L DIMENSIONS STAINLESS STEEL VERSION Compact Version 171 156 186 146...
Страница 12: ...11 English SE58 L Separate Version 236 40 7 90 241 55...