ALARM MESSAGES (CAUSES AND ACTIONS TO BE TAKEN)
MESSAGE
CAUSES
ACTION TO TAKE
NO ALARMS
All works regularly
---
[000] SYSTEM RESTART
---
---
[001] INTERNAL PS FAIL Internal supply voltage error
Contact the service
[002] CLOCK NOT SET
System Clock not set
Set the system clock from the transmitter
menu 13 (see also MCP function ).
[003] SD CARD FAILURE SD card not found or unreadable
check and/or replace SD card
[005] F-RAM ERROR
Error writing / reading Flash-RAM
Contact the service
[006] EXCITATION
ERROR
The excitation of the sensor coils
resulting from cable is interrupted
Check the connecting cables to the sensor.
[007] SIGNAL ERROR
The measure is strongly effected
by external noise or the cable
connecting the transmitter to the
sensor is broken.
Check the status of the cables connecting
the sensor, the grounding connections of the
devices and the possible presence of noise
sources.
[008] PIPE EMPTY
The measuring pipe is empty or
the detection system has not been
properly calibrated.
Check whether the pipe is empty or repeat
the empty pipe calibration procedure.
[009] FLOW>MAX+
The flow rate is higher than the
maximum positive threshold set.
Check the maximum positive flow rate
threshold set and the process conditions.
[010] FLOW>MAX-
The flow rate is higher than the
maximum negative threshold set.
Check the maximum negative flow rate
threshold set and the process conditions.
[011] FLOW<MIN+
The flow rate is lower than the
minimum positive threshold set.
Check the minimum positive flow rate
threshold set and the process conditions.
[012] FLOW<MIN-
The flow rate is lower than the
minimum negative threshold set.
Check the minimum negative flow rate
threshold set and the process conditions.
[013] FLOW>FULL
SCALE+
The flow rate is higher than the
full scale positive value set on the
device.
Check the full scale positive value set on the
device and the process conditions.
[014] FLOW>FULL
SCALE-
The flow rate is higher than the
full scale negative value set on the
device.
Check the full scale negative value set on the
device and the process conditions.
[015] PULSE1>RANGE
The pulse generation output 1 of
the device is saturated and cannot
generate the sufficient number of
impulses.
Set a bigger unit of volume or, if the
connected counting device allows it, reduce
the pulse duration value.
[016] PULSE2>RANGE
The pulse generation output 2 of
the device is saturated and cannot
generate the sufficient number of
impulses.
Set a bigger unit of volume or, if the
connected counting device allows it, reduce
the pulse duration value.
[017] CALIBR.ERROR
Calibration Error
Contact the service
[018] SYSTEM FREQ.
ERR
System Freq. Error
Contact the service
[019] B.DATA NOT INIT
Uninitialized data system
Contact the service
[020] FL.SENSOR
ERROR
Flow rate sensor error
Contact the service
[021] BATTERY LOW
(Rechargeable) battery depleted
Contact the service to Replace the battery
Содержание B0A1C2B1A4A0
Страница 1: ...Type SE58 L Transmitter Operating Instructions...
Страница 8: ...7 English SE58 L Dimensions 200 114 164 167 114 180 290 195...
Страница 9: ...8 English SE58 L Remote version 142 28 295 120 7 30...
Страница 11: ...10 English SE58 L DIMENSIONS STAINLESS STEEL VERSION Compact Version 171 156 186 146...
Страница 12: ...11 English SE58 L Separate Version 236 40 7 90 241 55...