64
Troubleshooting
Type 8745 Ethernet, 8745 Analogue
12.3.8
Unstable measured value (MFM)
You have not connected the functional earth (FE) properly.
→
To connect the functional earth, use a green-and-yellow cable
that is as short as possible. And the cable cross-section must
be at least equal to the cross section of the power supply
cable. Refer to chpt.
12.3.9
Unstable measured value (MFC)
Identify the cause to solve the problem:
1.
You have not connected the functional earth (FE) properly.
→
To connect the functional earth, use a green-and-yellow cable
that is as short as possible. And the cable cross-section must
be at least equal to the cross section of the power supply
cable. Refer to chpt.
2.
The residual ripple on the voltage supply is too high.
→
Use a supply voltage that conforms to the technical data
given in chpt. 6.9 Electrical data
.
3.
The product must compensate for irregularities in an unstable
pressure supply caused, for example, by pumps.
→
Install a suitable pressure regulator in front of the product.
→
Install a buffer tank to absorb the pressure fluctuations.
12.3.10 Set-point value at 0 %, but operating
fluid still flows (MFC)
The operating pressure is above the tight sealing pressure of the
control valve.
→
Reduce the operating pressure.
→
To eliminate the defect, return the product to the
manufacturer.
12.3.11
Set-point value at 0 %, control valve is
closed, no mass flow, but a non-zero
mass flow rate is measured (MFC)
Identify the cause to solve the problem:
1.
The installation position of the product is incorrect.
→
Install the product in the position indicated on the calibration
plate or in the calibration protocol.
→
Run the Autotune function to adapt to the operating
conditions.
2.
The operating fluid is different from the fluid specified during
the calibration.
→
Use the specified operating fluid or send the product to the
manufacturer for calibration with the new operating fluid.