5 Putting into operation
50
B-90
Operation Manual, Version B
Example photo with attached feeding tube and adapter piece (transparent adapter on the pump inlet at the right side).
Table 5-1: Choosing the feeding tube material
Solvent
Silicone tube
Tygon MH 2075
Tygon F 4040 A
Methanol
suitable
suitable
suitable
Ethanol
suitable
suitable
suitable
Acetone
unsuitable
suitable
unsuitable
Toluene
unsuitable
unsuitable
suitable
Isopropanol
suitable
suitable
suitable
Chloroform
unsuitable
unsuitable
unsuitable
Dichloromethane
unsuitable
unsuitable
unsuitable
THF
unsuitable
unsuitable
unsuitable
Ethylacetate
unsuitable
suitable
unsuitable
Hexane
unsuitable
unsuitable
suitable
Water
suitable
suitable
suitable