8 Troubleshooting
89
B-90
Operation Manual, Version B
Table 8-1: Error messages and their remedy (cont.)
043
O
2
sensor short-circuit
Defective sensor, sensor cable or
internal wiring
Contact BUCHI customer service
046
O
2
concentration high
System not leakproof, N
2
-bottle
empty
Test for leakages. Check system
sealings and N
2
-supply.
047
O
2
concentration low
Valve defective
Contact BUCHI customer service
050
B-295 O
2
concentration
too high
System not leakproof, N
2
-bottle
empty
Test for leakages. Check system
sealings and N
2
-supply.
051
Change gas mode
Electrical communication cable
B-290 / B-295 or dongle not
connected
Check cabling or dongle.
058
EEPROM
Contact BUCHI customer service.
Only for operation with the Inert Loop B-295
200
Connection to B-295 interrupted
or disconnected
Defective connection cable or
instrument is not plugged in
Check connection cable for
damage and insert plug.
201
B-295 plugged in while opera-
tion in process
Connection cable plugged in
after switching on the instrument
Restart the system.
203
Pressure alarm not given during
start-up
Pressure switch defective or
contaminated
Contact BUCHI customer service.
204
Failure cooling unit B-295
Contact BUCHI customer service.
8.2
Malfunctions and their remedy
Table 8-2: Malfunctions and their remedy
Malfunction
Possible cause
Remedy
Equipment cannot be switched on
No voltage
Insert mains plug, examine plug for
damage
Fuse has been activated
Reset fuse manually
Peristaltic pump does not deliver
The rollers are not in contact with the
running surface
Elevate the running surface with the
lever. Adjust the lifting height of the
running surface with the hexagonal
hex-wrench from beneath
System does not heat up
Heater cable not plugged in
Plug in heating circuit
Heating is not switched on
Switch on heat
Nominal entry temperature is below
room temperature
Re-select a new inlet temperature
Fuse has been activated
Reset fuse manually
Heater defective
Contact the BUCHI customer service
Faulty gas flow direction or no flow
in the heating system
Check pipe system