Büchi Labortechnik AG
Installation | 5
Operation Manual Lyovapor™ L-200/L-200 Pro
31/100
NOTICE
Contamination of the device
Escaping condensate can contaminate the instrument.
u
Fit the condensate drain hose with a downward slope. Make sure that the drain
hose is not immersed in the condensate.
u
Seal the condensate drain hose with the end plug as soon as cleaning of the
instrument is completed.
1
2
3
Fig. 13: Condensate drain hose
1
Rubber ring
2
Drain hose for condensate
3
End plug
The condensate drain hose is on the side of the instrument. The drain hose is sealed
with an end plug. The condensation is drained off through this drain hose after
completion of the freeze-drying process.
u
Pull the drain hose (2) together with end plug (3) out of the side panel of the
instrument.
u
Remove the end plug.
u
Direct the drain hose into a waste outlet or place a container underneath it.
5.4
Commissioning the vacuum pump
The vacuum pump evacuates the top-mount drying rack during the freeze-drying
process.