7 | Operating Pro control panel
Büchi Labortechnik AG
60/100
Operation Manual Lyovapor™ L-200/L-200 Pro
u
Tap the function
[Save]
on the function bar.
ð
The setting is changed.
7.7
Deleting a method
Navigation path
➔
Method
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The control panel highlights the selected method in green.
u
Tap the function
[Delete]
on the function bar.
u
Select
[Confirm]
to confirm the action in response to the confirmation question.
ð
The method is deleted.
7.8
Setting end point definitions
The end of a phase can be automatically set for each preparation by defining an end
point.
The end point can be defined by means of the temperature difference test or the
pressure difference test.
7.8.1
Connect the sensors for pressure difference test (optional feature)
u
Fit the pressure sensor, seal to the connection
and fix with the clamp.
u
Plug the pressure sensor electrical connector
into the socket marked Internal Vacuum
Sensor .
u
Fit the pressure sensor, seal to the connection
and fix with the clamp.
u
Plug the pressure sensor electrical connector
into the socket marked External Vacuum
Sensor.
u
On the control panel, select the sensor Inficon
Porter CDG020 D on the
[Settings]
submenu.