7 | Operating Pro control panel
Büchi Labortechnik AG
56/100
Operation Manual Lyovapor™ L-200/L-200 Pro
u
Enter the gas type.
u
Tap the function
[Save]
on the function bar.
ð
The setting is saved.
ð
The dialog box closes.
7.6.5
Setting the shelf loading temperature
Navigation path
➔
Method
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The control panel highlights the selected method in green.
u
Tap the action
[General]
.
ð
The control panel shows the action
[General]
.
u
Tap the action
[Shelf load temp.]
ð
The control panel shows a dialog box with a numeric input box.
u
Enter the value in the numeric input box.
u
Tap the function
[Save]
on the function bar.
ð
The setting is saved.
ð
The dialog box closes.
7.6.6
Setting the steps of a method
The Pro control panel can save up to 30 steps for each method.
NOTE
The maximum heating rate is 3 °C/min.
NOTE
The settings for the action Steps affect a single step in each case.
Navigation path
➔
Method
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The control panel highlights the selected method in green.
u
Tap the action
[Steps]
.
ð
The control panel shows the action Steps.
The following settings are available for each step:
Setting
Option
Meaning
[Step phase]
Primary drying/Sec-
ondary drying
Sets the type of step phase.
[Term]
Enter value
Sets the duration of the step in
minutes.