5 | Installation
Büchi Labortechnik AG
26/100
Operation Manual Lyovapor™ L-200/L-200 Pro
Precondition:
R
The electrical installation is as specified on the
type plate.
R
The electrical installation is equipped with a
proper grounding system.
R
The electrical installation is equipped with
suitable fuses and electrical safety features.
R
The installation site is as specified in the
technical date. See Chapter 3.6 "Technical
data", page 19
u
Connect the power supply cable to the
connection on the instrument. See Chapter 3.2
"Configuration", page 15
u
Connect the mains plug to an own mains outlet
socket.
5.3.3
Fitting the advanced vacuum control unit
u
Fit the 90° elbows onto the cross-pipe.
u
Fit the aeration valve onto the 90° elbow.
Precondition:
R
The arrow is pointing in the direction of the
vacuum pump.
u
Fit the main valve onto the cross-pipe.