7 | Operating Pro control panel
Büchi Labortechnik AG
58/100
Operation Manual Lyovapor™ L-200/L-200 Pro
Navigation path
➔
Method
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The control panel highlights the selected method in green.
u
Tap the action
[Phase]
.
ð
The control panel shows the Phase view.
The following phases of a method are available:
Phase
Setting
Option
Meaning
[Primary
drying]
[Pressure ac-
tion]
None/Sample
protection/
Message
None: no action is carried out.
Sample protection: if the pressure is
too high, heating of the shelves is
paused.
Message: if the pressure is too high,
the control panel displays a message.
[Temp. action]
None/Sample
protection/
Message
None: no action is carried out.
Sample protection: if the tempera-
ture is too high, heating of the
shelves is paused.
Message: if the temperature is too
high, the control panel displays a
message.
[Safety temper-
ature]
Enter value
Maximum divergence from the set
collapse temperature before the
sample protection function is acti-
vated.
[Safety temp.
Time]
Enter value
Time as of which sample protection
is inactive. The value relates to the
time before completion of primary
drying.
[End point defi-
nition]
More settings