BÜCHI Labortechnik AG
Performing an extraction | 9
Operation Manual FatExtractor E-500
63/88
u
Put the glass sample tube with frit into the
holder in the extraction chamber.
9.3
Performing a Soxhlet Extraction
9.3.1
Adjusting the optical sensor on the extraction glass chamber Soxhlet
u
Adjusting the optical sensor on the extraction
glass chamber Soxhlet.
White line to the center of the upper sand layer.
Sand
Celite®
+ sample
Sand
9.3.2
Preparing the instrument for a Soxhlet Extraction
Precondition:
R
All commissioning operations have been completed. See Chapter 5 "Installation",
page 27
R
Make sure that the configuration Soxhlet Extraction is selected. See Chapter 5.14
"Selecting a configuration", page 39
R
Make sure that there are no extraction beakers in the instrument.
R
Make sure that no defective sealings or glass parts are used.
R
u
Switch the
On/Off
master switch to On.
ð
The instrument is starting up.
9.3.3
Selecting a method
Navigation path
➔
u
Navigate to the
[Home]
menu via the navigation path.
u
Tap the method.
ð
The display shows a dialog with the selectable methods.